Measured Components SO2, NO, NO2, NH3 Measurement Principles Differential Optical Absorption Spectroscopy (DOAS) Measuring Ranges
SO2: 0~50 / 0~100 / 0~200 / 0~500 / 0~1,000 / 0~2,000 ppm NOx: 0~40 / 0~100 / 0~150 / 0~200 / 0~500 / 0~1,000 / 0~2,000 ppm
NH3: 0~10 / 0~20 / 0~50 ppm Min. Measuring Unit 0.1ppm Accuracy NH3: < ±1.0 ppm / NO: < ±0.5 ppm
NO2: < ±2.5 ppm / SO2: < ±0.5 ppmZero Drift (24 hours) 〈 ± 1% FS Span Drift (24 hours) 〈 ± 1% FS Repeatability 〈 ± 1% FS Linearity
〈 ± 1% FS Response Time < 5 seconds Certificates / Approvals
-
Type Approval (Korea)
Performance Approval (Korea)
CPA, CCEP (China)
TUV (Germany)
-
-
System Components
-
DGA-X Main Unit / Probe / Power Distribution Panel
/ Air Purge Unit / Master Flange / Cables
DGA - X Features
-
Simultaneous measurements for two components among NOx, SO2, and NH3
Less interenference from moisture and particles as it uses UV light source
NOx converter is not required as it measures NO and NO2 separately
Enclosure level up to IP66 and NEMA4X
Real Gas Calibration using patented Auto Calibration Unit (ACU)
EX version is available as an option
Options
-
Probe Protector / Teflon Coated Probe
ACU (Auto Calibration Unit)
Higher Enclosure or Protection Level
Regulator & Valve / Calibration Gas
-
-
Measurement Conditions
Operating Temperature -20 ~ +60 ℃ Gas Temperature < +900 °C Gas Pressure ±60 hPa (±611 mmH20) Communication
Analog Outputs 2 Channel, 4 ~ 20 mA Digital Outputs 4 Channel Digital Inputs 2 Channel Display & Input Device 7" LCD Monitor(Touch screen) Interface RS232,422,485 / LAN / Hart -
Dimension & Power Supply
Dimensions W300 x D420 x H413 mm Weight 22 kg Enclosure Rating P65 (IP66, NEMA4X) Voltage 100 ~240 VAC, 50/60 Hz Power Consumption 500 W Probe
Materials SUS 316L or SUS 316Ti Length 1.0 M, 1.5 M, 2.0 M (Adjustable) Measurement Section Length 300 mm, 500 mm other length available (Optional) Gas Flow Rate > 1 m /s Weight 1.5 M : 20 kg 2.0 M : 25 kg Probe adapter : 5 kg Air Purge Necessary Temperature Sensor PT 1000