Measured Components CO Measurement Principles TDLS (Tuneable Diode Laser Spectroscopy)
Measuring Ranges
CO : Min 0~1000 ppm Accuracy < ±1.0 ppm Zero Drift (24 hours) 〈 ± 1% FS
Span Drift (24 hours) 〈 ± 1% FS
Repeatability 〈 ± 2% FS
Linearity
〈 ± 1% FS
Response Time < 5 seconds -
System Components
-
RGA-60 Main Unit / Probe / Power Distribution Panel
Air Purge Unit / Master Flange / Cables
RGA - 60 Features
-
No interference from other gas components by TDLS measuring principle
Less interference from moisture and dust
Rapid response time and high sensitivity by using laser scanning of extremely narrow range
Options
-
Probe Protector / Teflon Coated Probe
Higher Enclosure or Protection Level
Regulator & Valve / Calibration Gas
-
-
Measurement Conditions
Operating Temperature -20 ~ +55 ℃ Gas Temperature < +450 °C Communication
Analog Outputs 2 Channel, 4 ~ 20 mA Digital Outputs 4 Channel Digital Inputs 2 Channel Display & Input Device 7" LCD Monitor(Touch screen) Interface RS232,422,485 / LAN / Hart -
Dimension & Power Supply
Dimensions W400 x D201 x H500 mm Weight 18.5 kg Voltage 100 ~240 VAC, 50/60 Hz Power Consumption 200 W (Max.) Probe
Materials SUS 316L or SUS 316Ti Length 0.5 M ~ 2.5 M (Adjustable) Measurement section Length 500 mm Gas Flow Rate > 1 m /s Weight 1.5 m : 20 kg
2.0 m : 25kg
Temperature Sensor PT 1000