Measured ComponentsHCl Measurement Principles TDLS (Tunable Diode Laser Spectroscopy) Measuring Ranges
Min 0~20 / Max 0~100 ppm Accuracy < ±2% FS
Zero Drift (24 hours) 〈 ± 2% FS
Span Drift (24 hours) 〈 ± 2% FS
Repeatability 〈 ± 2% FS
Linearity
〈 ± 2% FS
-
System Components
-
LSM-30 Main Unit
Features
-
Less interference from other other gascomponents as TDL scans very narrow range of light wavelength (0.035 um)
Less interference from moisture and dustcompared to other measuring principles
Analysis in low flow rate is possible by minimizing measuring cell capacity
Options
-
Sample Probe / Sample Line / Heating Block
Distribution Panel
Rack Panel / Regulator / Calibration Gas
-
-
Measurement Conditions
Operating Temperature -20 ℃ ~ +50 ℃ (-40 ℃ ~ +80 ℃) Operating Humidity 0 ~ 99% r.H. Gas Temperature < +190 °C Cell Temperature +190 °C Cell Heating time 45 min (When heating from +25°C) Gas Humidity
MAX 20% abs. H20 Gas Flow Rate 1.0 L/min ~ 5.0 L/min Sampling Pump Heated Diaphragm pump Sampling Tube PTFE Operating Pressure 800 ~ 1,100 mbar Communication
Analog Outputs 2Channel, 4 ~ 20 mA Digital Outputs
12 VDC, 4 Channel Digital Inputs
Digital 4Ch / Current, DI Voltage 12 VDC ~ 48 VDC Display & Input Device 7" LCD Monitor (Touch screen / USB) Interface RS232,422,485 / TCP-IP Via Ethernet / Hart Communication -
Dimension & Power Supply
Materials SUS 304, Al 6061 Dimensions W440 x D550 x H240 mm Weight 20 kg
Voltage 88 ~ 264 VAC, 48 ~ 63 Hz Power Consumption MAX 165 W